Welch Allyn Connex VSM 6400 Vital Signs Monitor
The Welch Allyn Connex VSM 6400 Vital Signs Monitor is an advanced, touch screen monitor featuring bright, vivid colors with minimal knobs and buttons for ease of cleaningโperfect for hospital general care floors and for surgery centers and other ambulatory care offices. The Connex VSM helps clinicians dramatically improve patient safety and the time it takes for vitals capture. A wide range of optional configurations help clinicians to reduce/improve common challenges including falls, pressure ulcers, respiratory distress, EMR documentation, early warnings/detection, white coat hypertension and accuracy of vitals information.
Welch Allyn Connex VSM 6400 Vital Signs Monitor Features
- SureBPยฎ supports all ages and provides 15-second readings as the cuff inflates
- SureTempยฎ Plus and/or Braun ThermoScanยฎ PRO 6000 Ear Thermometry
- Single sign-on options available for quick, secure clinician login
- Masimoยฎ or Nellcorยฎ Pulse Oximetry (SpO2) Algorithms
- Optional 3- or 5-Lead ECG bedside monitoring
- Optional Masimoยฎ noninvasive continuous monitoring of total hemoglobin (SpHb)
- Optional Covidienยฎ Capnography (CO2)
- Optional custom data inputs
- Optional Masimoยฎ Acoustic Respiratory Monitoring (RRa)
- Optional early warning score integration
- Health o meterยฎ, Seca and Detecto weight scale connectivity
- Wireless connectivity to leading electronic medical records
- Optional EarlySenseยฎ contact-free respiratory monitoring and motion sensing
Welch Allyn Connex VSM 6400 Vital Signs Monitor Specifications
-
Electrical rating: 100 – 240 VAC | 50 – 60 Hz | 0.8 – 1.5 A
-
Duty cycle: Continuous operation
-
Type of protection against electric shock: Class I
- Dimensions: 11″ W x 6″ D x 10″ H
- Weight: 10 lbs
-
Display area: 8″ H x 4″ V
-
Pixels: 1024 (H) x 600 (V)
-
Pixel arrangement: RGB
-
Color depth: 16 bits per pixel
-
Output sound pressure: 67 dB at 1.0 yard
-
Pulse frequency: 150 – 1000 Hz
-
Number of harmonic components in the range 300 Hz to 4000 Hz: minimum of 4
-
Effective pulse duration
-
high priority: 75 – 200 ms
-
medium and low priority: 125 – 250 ms
-
NIBP
-
Cuff pressure range: Meets or exceeds ANSI/AAMI SP10:2002 standards for cuff pressure range
-
Systolic range
-
Adult: 30 to 260 mmHg (4.0 to 34.7 kPa) (StepBP, SureBP)
-
Pediatric: 30 to 260 mmHg (4.0 to 34.7 kPa) (StepBP, SureBP)
-
Neonate: 20 to 120 mmHg (2.7 to 16.0 kPa) (StepBP)
-
- Diastolic range
- Adult: 20 to 220 mmHg (2.7 to 29.3 kPa) (StepBP, SureBP)
- Pediatric: 20 to 220 mmHg (2.7 to 29.3 kPa) (StepBP, SureBP)
- Neonate: 10 to 110 mmHg (1.3 to 14.7 kPa) (StepBP)
- Cuff Inflation Target
- Adult: 160 mmHg (21.3 kPa) (StepBP)
- Pediatric: 120 mmHg (16.0 kPa) (StepBP)
- Neonate: 90 mmHg (12.0 kPa) (StepBP)
- Maximum Target Pressure
- Adult: 280 mmHg (37.3 kPa) (StepBP, SureBP)
-
Pediatric: 280 mmHg (37.3 kPa) (StepBP, SureBP)
-
Neonate: 130 mmHg (17.3 kPa) (StepBP)
- Blood pressure determination time
- Typical: 15 seconds
- Maximum: 150 seconds
- Blood pressure accuracy: Meets or exceeds ANSI/AAMI SP10:2002 standards for noninvasive blood pressure accuracy, ยฑ5 mmHg (0.7 kPa) mean error, 8 mmHg (1.1 kPa) standard deviation
- Mean Arterial Pressure (MAP) range
- Adult: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
- Pediatric: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
- Neonate: 13 to 110 mmHg (1.7 to 14.7 kPa) (StepBP)
- Pulse rate range (using blood pressure determination)
- Adult: 30 to 200 beats per minute (StepBP, SureBP)
- Pediatric: 30 to 200 beats per minute (StepBP, SureBP)
- Neonate: 35 to 220 beats per minute (StepBP)
- Pulse rate accuracy (using blood pressure determination): ยฑ5.0% (ยฑ3 beats per minute)
- Overpressure cutoff
- Adult: 300 mmHg ยฑ15 mmHg (40.0 kPa ยฑ2.0 kPa)
- Pediatric: 300 mmHg ยฑ15 mmHg (40.0 kPa ยฑ2.0 kPa)
- Neonate: 150 mmHg (20.0 kPa) maximum
Reviews